Application of Ultrafiltration in High-concentration Formulations Production

2024.02.18 864


For antibody drugs, what is usually referred to as high-concentration formulations are formulation products with a final antibody concentration of >100mg/ml. Currently, about one-third of FDA-approved monoclonal antibody products are high-concentration formulations (>100mg/ml), of which close to 80% were approved after 2015, which shows that there is an increasing demand for high-concentration formulations in the biological industry.


Advantages and Challenges of High-concentrated Formulations

Advantages

The advantages of high-concentration formulations are self-evident, especially in subcutaneous injections, which require a concentration of 100mg/ml or more due to the limitation of the volume of subcutaneous injections.


Challenges

The challenges associated with high-concentration formulations include increased viscosity of the material, volume exclusion issues, the Donnan effect, system hardware design and scale-up issues.


1. Increased viscosity of the material

As the concentration increases, interactions between protein molecules are enhanced which results in viscosity rises, mobility deteriorates, and severe flux attenuation.


2. Volume exclusion issues

At high concentrations, the permeate concentration of relatively small solute components upstream and downstream of the membrane changes, leading to an increase in permeate pressure.


3. The Donnan effect

Charged solutions are trapped by the ultrafiltration membrane, and charge interactions result in uneven distribution of the charged solutes on both sides of the membrane, leading to changes in concentration and pH.


4. System hardware design and scale-up issues

During the scale-up process, the system needs to be designed with a minimum working volume in mind to avoid sample loss.

There are improvements in structural design in Cobetter's conventional screen ultrafiltration TFF cassettes, so they have certain advantages in high-concentration formulation production.


Examples

The following are examples of Cobetter's conventional screen TFF cassettes in the application of high-concentration formulations:


Case 1

Lab scale, monoclonal antibody product


TFF Cassette Format: UFCLA0030010P (RC, 30 kDa, 0.11m2)


Sample information: Monoclonal antibody product, the initial concentration of the solution is 18g/L, and the inlet flow rate is 550ml/min, the performance is shown in the table below.

Table 1 Test Results


The trend of inlet pressure and flux with time during the experiment is shown in the figure below:


Figure 1 Trends in Inlet Pressure And Flux through Time

In the late stage of the experiment, with the increase in concentration, the viscosity of the solution increased exponentially, the inlet pressure increased obviously, and it was necessary to continuously reduce the inlet flow rate to control the pressure increase. The experiment was stopped when the inlet flow rate was reduced to 1/6 of the initial flow rate, and the sample recovery reached 98% after flushing with the buffer which is 2 times the working volume. After the experiment, washing with 0.1 M NaOH for 30 min, the water flux of the cassette was recovered by 92% before and after use.

Case 2

Lab scale, monoclonal antibody product


TFF Cassette Format: UFCLA0030001P (RC, 30 kDa, 0.01m2)

Sample information: Monoclonal antibody product, the initial concentration of the solution is 21g/L, and the inlet flow rate is 55ml/min, the performance is shown in the table below:

Table 2 Test Results

The trend of inlet pressure and flux with time during the experiment is shown in the figure below:


Figure 2 Trends in Inlet Pressure and Flux through Time

During the experiment, the pressure change was obvious, and it was necessary to constantly adjust the retentate valve. Stopped the experiment after the concentration factor had reached the requirement, then flushing with the buffer that is 2 times the working volume. The sample recovery rate reached 95%. After the experiment, use 0.1M NaOH to clean for 30 min, and the water flux of the TFF cassette recovered 93%.

Case 3

Pilot scale, monoclonal antibody product


TFF Cassette Format: UFCFL0030250P (RC, 30kDa, 2.33m2)

Sample information: Monoclonal antibody product, the initial concentration of the solution is 14g/L, and the inlet flow rate is 699ml/min, the performance is shown in the table below:

Table 3 Test Results

The sample recovery rate reached 93% after flushing with the buffer that is 2 times the working volume and emptying the air. After the experiment, used 0.1M NaOH to clean for 30 min, and the water flux of the TFF cassette resulted in 100% recovery before and after use.

Cobetter catches up with the market demand, constantly improves the products, and always be committed to providing customers with high-quality and trustworthy products, helping customers to reduce costs and improve the safety of drug production. If you want to know more about our products or experimental requirements, please contact [email protected].



You may also be interested in

夜夜黄片| 在线毛片| 欧美福利片一级| 亚洲色情18在线观看| 欧美日韩国产的0| 人妻少妇妇一区二区| www夜夜AV| 国产性色在线| www.久久艹| 欧美日韩精品亚洲综合久久| 亚洲熟女三区四区| 欧美在线观看一区| 午夜按摩视频在线观看视频| 欧美激情在线视频免费视频| 香蕉av第一页| 精品美女久久| http久久视频| 污秽一区二区三区中文| 美国无码一区二| 美女推油一区二区三区| 日韩精品在线观看免费| AV播放国产| 日本第一福利成人官网| 韩国日本三级一区二区三区| 老司机高清无码中文字幕| 肛交伊人视频| 肏逼视频91免费版| 嗯嗯啊啊嗯视频免费| 久久久久久久久久久久久久蜜臀 | 992午夜福利姬在线观看| 青青青青久久国产精品| 嗯嗯啊嗯啊嗯啊~视频| 久久久久无码精品国产高潮| 伊人AV二区| 人人人美女精品视频在线观看| 加勒比一二三四在线| www啪啪啪日本com| 勉费搞B视屏| 二区三区国产精品| 黄片在线免费观看| 欧美 国产一区二区|